Resumen
The clinical efficiency of a photo polymerization unit is
crucial to obtain an optimal polymerization and a successful result. The present
observational study aimed to evaluate and analyze the intensity of the luminous power of the
photo polymerization lamps of private dental clinics in Cuenca City, Ecuador.The luminous
intensity was evaluated using the CK12024 radiometer (measured the curing time up to2000).
It was classified according to the following parameters: Low performance
(<300mw/cm2),regular performance (300 – 800 mw/cm2) and high performance (>800
mw/cm2), in a time of 20seconds and 40 seconds. in “Microsoft Excel”,;to register all the
necessary data: number of lamp, typeof lamp, intensity of light power in 20 seconds,
intensity of light power in 40 seconds, breakage of optical fiber and residues of
biomaterials in the optical fiber. If you obtained as a result that of the 58lamps
evaluated, a percentage of 0.6% (n=1) of the total photo polymerization units had a low
performance, a 5.2% (n=3) had a medium performance and a 93.1% (n = 54) of them had a high
performance. An adequate polymerization is crucial to obtain good physical and mechanical
properties and excellent clinical results. According to the results obtained within the
present study, it is observed that the majority of the photo polymerization lamps used in
the private offices of the City of Cuenca, have a high clinical performance.
| Idioma original | Español |
|---|---|
| Publicación | RECISATEC – Revista científica Saúde e tecnologia |
| Estado | Publicada - 2023 |
| Publicado de forma externa | Sí |
Palabras clave
- Intensidad lumínica; Lámparas de fotopolimerización; Rendimiento clínico